Agilent Technologies, Inc. E1529A
Stock photo

Agilent Technologies, Inc. E1529A Remote Strain Conditioning Unit

Feature:
Type: Remote Strain Conditioning Unit
Size: C

The HP E1529A remote strain conditioning unit simplifies structural testing of designs that require many channels of strain measurement. Strain gage cables connect to the unit with standard RJ-45 telecom plugs, making it exceptionally easy to reconfigure or replace strain gages.

Each channel has bridge-completion and shunt-calibration resistors (with provision for special values). Automatic resetting of fuses ensures that a shorted gage does not disturb any other channel. Up to four external sources can be used to generate the excitation voltages for banks of eight gages.

The HP E1529A provides gain, filtering and 32:1 multiplexing. It connects to a C-size VXI remote-channel DAC module (the HP E1422A), which provides setup and configuration of up to 16 HP E1529A units. The HP E1422A controls the HP E1529A multiplexer and includes a 16-bit, 25 kilosample/s scanning digitizer for signal acquisition. The connections and I/O control of E1529A's is provided by HP E1529A remot
32 Channels of Conditioning
Outputs For Static-Load or Vibration
Locate Up to 1000 Ft. From Test System
Internal Self-Calibration Source
32:1 Multiplexing
Connects to DAC module (E1422A)
Rocket & Satellite Structural

 